Don't show this message again
§ 84.304 Capacity test requirements.
(a) NIOSH will conduct the capacity test on a total of 8 to 10 of the units submitted for approval, as follows:
(1) Three units will be tested on a breathing and metabolic simulator in the condition in which they are received from the applicant;
(2) Two units will be tested on a breathing and metabolic simulator after being subjected to the environmental treatments specified in § 84.307 of this subpart;
(3) Two units will be tested on a breathing and metabolic simulator at the cold-temperature limit recommended by the manufacturer under § 84.302(h)(1), after the unit has been stored for a minimum of 24 hours at this limit; and