§ 86.522-78 Carbon monoxide analyzer calibration.
(a) Initial and periodic interference check. Prior to its introduction into service and annually thereafter the NDIR carbon monoxide analyzer shall be checked for response to water vapor and CO2:
(1) Follow the manufacturer's instructions for instrument startup and operation. Adjust the analyzer to optimize performance on the most sensitive range.
(2) Zero the carbon monoxide analyzer with either zero grade air or zero grade nitrogen.