§ 1065.590 PM sampling media (e.g., filters) preconditioning and tare weighing.
Before an emission test, take the following steps to prepare PM sampling media (e.g., filters) and equipment for PM measurements:
(a) Make sure the balance and PM-stabilization environments meet the periodic verifications in § 1065.390.
(b) Visually inspect unused sample media (e.g., filters) for defects and discard defective media.
(c) To handle PM sampling media (e.g., filters), use electrically grounded tweezers or a grounding strap, as described in § 1065.190.
(d) Place unused sample media (e.g., filters) in one or more containers that are open to the PM-stabilization environment. If you are using filters, you may place them in the bottom half of a filter cassette.
(e) Stabilize sample media (e.g., filters) in the PM-stabilization environment. Consider an unused sample medium stabilized as long as it has been in the PM-stabilization environment for a minimum of 30 min, during which the PM-stabilization environment has been within the specifications of § 1065.190.
(f) Weigh the sample media (e.g., filters) automatically or manually, as follows: