NO2-to-NO converter conversion verification.

§ 1065.378 NO2-to-NO converter conversion verification.

(a) Scope and frequency. If you use an analyzer that measures only NO to determine NOX, you must use an NO2-to-NO converter upstream of the analyzer. Perform this verification after installing the converter, after major maintenance and within 35 days before an emission test. This verification must be repeated at this frequency to verify that the catalytic activity of the NO2-to-NO converter has not deteriorated.

(b) Measurement principles. An NO2-to-NO converter allows an analyzer that measures only NO to determine total NOX by converting the NO2 in exhaust to NO.

(c) System requirements. An NO2-to-NO converter must allow for measuring at least 95% of the total NO2 at the maximum expected concentration of NO2.

(d) Procedure. Use the following procedure to verify the performance of a NO2-to-NO converter:

(1) Instrument setup. Follow the analyzer and NO2-to-NO converter manufacturers' start-up and operating instructions. Adjust the analyzer and converter as needed to optimize performance.